Tubing, Loops, Columns, and Fittings
Instrument Guide | 7 7
Sample Injection Loops
Columns
The NGC chromatography systems support columns for various chromatography
techniques. ChromLab includes a library of the most common commercial columns
in the Method Editor.
Table 6. NGC system injection loop details
Injection Loop Size Scope of Use Maximum Pressure
Rating
100 μl Small sample sizes 4,000 psi
250 μl Small sample sizes 4,000 psi
500 μl Small sample sizes 4,000 psi
1 ml Large sample sizes 3,000 psi
2 ml Large sample sizes 3,000 psi
5 ml Large sample sizes 3,000 psi
DynaLoop™ 25:
25 ml
Very large sample sizes 1,000 psi
DynaLoop 90:
90 ml
Very large sample sizes 1,000 psi