Using the C-96 Autosampler with NGC Systems
Instrument Guide | 293
To prime and purge the C-96 autosampler
Note: Before priming the autosampler, ensure that any column is offline.
1. Immerse the inlet tubing in a container of buffer or HPLC grade (filtered,
degassed) or other high quality water.
2. Use a syringe to prime the pumps. See the section Priming and Purging the
Systems on page 118 for specific information.
3. On the ChromLab™ computer or the touch screen, start ChromLab in Manual
mode.
4. Set the fluidic scheme to display the autosampler and sample inject valve.
5. In the fluidic scheme, touch or click the Sample Inject Valve module to open its
dialog box.
6. Set the inject valve to System Pump Inject Loop.
Autosampler
Sample Inject
valve
Vedere la pagina 294