A | Maintaining the Instrument
188 | NGC Chromatography Systems and ChromLab Software
Disassembling the Mixer
Use this procedure to replace any of the components of the mixer.
Important: Before disassembling the mixer, flush any hazardous material from
the NGC system. Ensure that the pumps are not running and bleed residual
pressure from the system.
To disassemble the mixer
1. On the touch screen, select Shut Down on the dropdown menu to exit
ChromLab and shut down the NGC instrument.
2. Disconnect the tubing from the top, bottom, and sides of the mixer.
Vedere la pagina 189