Plumbing NGC Systems
Instrument Guide | 283
Plumbing the NGC Discover Pro System
Table 23 lists the tasks for plumbing the NGC Discover Pro system. The tasks are
explained in detail in the sections that follow.
Note: Ensur
e that you complete the plumbing tasks in the sections Plumbing
All NGC Systems on page 265 and Plumbing the NGC Discover System on
page 278 before continuing.
The figure on page 284 shows the Point-to-Plumb connections for the
NGC Discover Pro system.
Table 23. Plumbing tasks for the NGC Discovery Pro systems
Tasks
1.
Tube #6 between the outlet port on the sample inlet valve and the
inlet port on the sample pump
Note: Remove the inlet tubing from the sample pump inlet
manifold if it is installed.
2.
Connect one of the following:
Tube #4 from the Common inlet port on the BioFrac fraction
collector diverter valve to port 1 on the outlet valve
Waste tubing to port 1 on the outlet valve
3.
Inlet tubing to the remaining ports on the sample inlet valve
See the section Plumbing Sample Inlet Valves on page 285.
4.
Outlet tubing to the remaining ports on the outlet valve
See the section Plumbing Outlet Valves on page 286.
Vedere la pagina 284