Bio-rad ChromLab™ Software Manuale Utente

Navigare online o scaricare Manuale Utente per Strumenti di misura Bio-rad ChromLab™ Software. Bio-Rad ChromLab™ Software User Manual Manuale Utente

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 306
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti

Sommario

Pagina 1 - Instrument Guide

NGC™ Chromatography Systems and ChromLab™ SoftwareInstrument GuideVersion 3.03.0

Pagina 2

viii | NGC Chromatography Systems and ChromLab SoftwareTable of ContentsOther Components . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 3 - Systems and

3 | Preparing the Instrument98 | NGC Chromatography Systems and ChromLab SoftwareChanging the Mixer BarrelThe mixer module is used to blend

Pagina 4

Changing the Mixer BarrelInstrument Guide | 9 9To insert or change the mixer barrel To insert or change the mixer barrel, see Replacing the Mixer

Pagina 5

3 | Preparing the Instrument100 | NGC Chromatography Systems and ChromLab SoftwareConnecting the Air Sensor ModuleThe air sensors enable the

Pagina 6

Connecting the Air Sensor ModuleInstrument Guide | 101Air Sensor Tubing Length and Bubble SizeThe response of the air sensor module is highly depe

Pagina 7 - Instrument Guide

3 | Preparing the Instrument102 | NGC Chromatography Systems and ChromLab SoftwareAttaching and Activating Air SensorsThis section explains

Pagina 8

Connecting the Air Sensor ModuleInstrument Guide | 1035. To connect an air sensor to the system pumps:a. Connect tubing from the buffer bottle or

Pagina 9 - Instrument Guide

3 | Preparing the Instrument104 | NGC Chromatography Systems and ChromLab SoftwareConnecting External Devices to the NGC InstrumentTip: If

Pagina 10 - Table of Contents

Connecting External Devices to the NGC InstrumentInstrument Guide | 105Importing Analog Signals to the NGC InstrumentYou can import analog (voltag

Pagina 11 - Instrument Guide

3 | Preparing the Instrument106 | NGC Chromatography Systems and ChromLab Software3. Insert the other end of the cable into one of the perip

Pagina 12

Connecting External Devices to the NGC InstrumentInstrument Guide | 107Connecting External Devices to the SIMWARNING! These procedures require y

Pagina 13 - 1 Introduction

Instrument Guide | i xTable of ContentsAppendix D Connecting the C-96 Autosampler to NGC Systems . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 14 - Plus system)

3 | Preparing the Instrument108 | NGC Chromatography Systems and ChromLab Software3. Loosen all three screws on the top of the connector.Tip

Pagina 15 - Main Features

Connecting External Devices to the NGC InstrumentInstrument Guide | 1099. On the analog output cable, determine the positive and negative wires.10

Pagina 16 - Site Requirements

3 | Preparing the Instrument110 | NGC Chromatography Systems and ChromLab Software11. Insert the negative wire and one end of the jumper wir

Pagina 17 - Bench Space

Connecting External Devices to the NGC InstrumentInstrument Guide | 11113. Insert the connector into the SIM 1 or SIM 2 port on the back of the SI

Pagina 18 - Safety Requirements

3 | Preparing the Instrument112 | NGC Chromatography Systems and ChromLab SoftwareTo connect an external device to the DAC A and DIGITAL OUT

Pagina 19 - Finding Out More

Connecting External Devices to the NGC InstrumentInstrument Guide | 1139. Insert the positive and ground wires into the appropriate terminals on t

Pagina 20 - 1

3 | Preparing the Instrument114 | NGC Chromatography Systems and ChromLab SoftwareStarting the NGC InstrumentCaution: The NGC system runs u

Pagina 21 - 2 The NGC Instrument

Starting the NGC InstrumentInstrument Guide | 115Connecting the NGC System to ChromLab SoftwareNote: For information about installing ChromLab, s

Pagina 22 - 2

3 | Preparing the Instrument116 | NGC Chromatography Systems and ChromLab SoftwareVerifying Plumbing with the Point-to-Plumb FeatureThe Poin

Pagina 23 - Instrument Guide

Verifying Plumbing with the Point-to-Plumb FeatureInstrument Guide | 117To verify the plumbing path 1. With both the Point-to-Plumb screen and the

Pagina 24 - Front View

x | NGC Chromatography Systems and ChromLab SoftwareTable of Contents

Pagina 25 - Instrument Guide

3 | Preparing the Instrument118 | NGC Chromatography Systems and ChromLab SoftwarePriming and Purging the SystemsThis section explains how t

Pagina 26 - Back View

Priming and Purging the SystemsInstrument Guide | 119Priming and Purging the NGC Quest SystemCaution: Before priming or purging the system, ensur

Pagina 27 - Right Side View

3 | Preparing the Instrument120 | NGC Chromatography Systems and ChromLab Software6. Remove the syringe from the priming port and discard th

Pagina 28 - Left Side View

Priming and Purging the SystemsInstrument Guide | 1215. In the fluidic scheme, touch or click the System Pump module to open its dialog box.6. In

Pagina 29

3 | Preparing the Instrument122 | NGC Chromatography Systems and ChromLab SoftwarePriming and Purging the NGC Scout SystemPriming the NGC Sc

Pagina 30

Priming and Purging the SystemsInstrument Guide | 1235. In the System Pump dialog box, select Priming on the Mode dropdown list.Note: When in Pri

Pagina 31 - System Pumps

3 | Preparing the Instrument124 | NGC Chromatography Systems and ChromLab SoftwarePurging the NGC Scout SystemTo purge the buffer blending v

Pagina 32 - How the System Pumps Function

Priming and Purging the SystemsInstrument Guide | 1253. To purge all lines of air bubbles and fill them with buffer:a. Click All.b. (Optional) Cha

Pagina 33 - Pump Priming Port

3 | Preparing the Instrument126 | NGC Chromatography Systems and ChromLab SoftwarePriming and Purging the NGC Discover SystemPriming the NGC

Pagina 34 - Piston Wash System

Priming and Purging the SystemsInstrument Guide | 1274. In the fluidic scheme, touch or click the System Pump module to open its dialog box.5. In

Pagina 35 - Sample Pump

Instrument Guide | 1 11 IntroductionThe NGC™ chromatography systems are preparative systems designed to rapidly automate the purification of biomo

Pagina 36 - How the Sample Pump Works

3 | Preparing the Instrument128 | NGC Chromatography Systems and ChromLab Software7. Insert a 30 ml syringe into the priming port on system

Pagina 37 - Instrument Guide

Priming and Purging the SystemsInstrument Guide | 129To prime the buffer inlet valves1. Immerse the tubing for the inlet valves in their respectiv

Pagina 38

3 | Preparing the Instrument130 | NGC Chromatography Systems and ChromLab Software8. Remove the syringe from the priming port and discard th

Pagina 39 - Sample Inject Valve

Priming and Purging the SystemsInstrument Guide | 131To prime the sample inlet valve (if available)1. Immerse the tubing for the inlet valve in th

Pagina 40 - Port Function

3 | Preparing the Instrument132 | NGC Chromatography Systems and ChromLab Software5. Insert a 30 ml syringe into the sample pump’s priming p

Pagina 41 - Sample Inject Valve LEDs

Priming and Purging the SystemsInstrument Guide | 133Purging the NGC Discover SystemTo purge the buffer blending valve1. Open the System Pump dial

Pagina 42 - Valve Position Explanation

3 | Preparing the Instrument134 | NGC Chromatography Systems and ChromLab Software2. To purge an individual line of air bubbles and fill it

Pagina 43

Priming and Purging the SystemsInstrument Guide | 135To purge the buffer inlet valves1. Open the System Pump dialog box and select Priming on the

Pagina 44

3 | Preparing the Instrument136 | NGC Chromatography Systems and ChromLab SoftwareTo purge the sample pump1. In the fluidic scheme, touch or

Pagina 45

Priming and Purging the SystemsInstrument Guide | 1373. Click Change Position to set the sample inject valve to Sample Pump Waste and close the di

Pagina 46 - Buffer Blending Valve LEDs

1 | Introduction12 | NGC Chromatography Systems and ChromLab SoftwareThe NGC Chromatography SystemsThe NGC chromatography systems are availa

Pagina 47 - Gradient Mode

3 | Preparing the Instrument138 | NGC Chromatography Systems and ChromLab Software2. In the Sample Pump dialog box, do one of the following:

Pagina 48 - Inlet Valves

Preparing the System and Sample Pump Piston Washing SystemsInstrument Guide | 139Preparing the System and Sample Pump Piston Washing SystemsNote:

Pagina 49 - Inlet Valve

3 | Preparing the Instrument140 | NGC Chromatography Systems and ChromLab SoftwarePreparing the ColumnsBefore a method can be run, the colum

Pagina 50 - Ports on the Inlet Valves

Preparing the ColumnsInstrument Guide | 141Connecting Column Clamps to the NGC InstrumentThe NGC instrument has indented metallic strips on both t

Pagina 51 - Inlet Valve LEDs

3 | Preparing the Instrument142 | NGC Chromatography Systems and ChromLab SoftwareConnecting Tubing to the ColumnsNote: Ensure that the sys

Pagina 52 - How the Inlet Valves Work

Preparing the ColumnsInstrument Guide | 143Washing the ColumnsCleaning a column requires access to ChromLab. You can perform this task from the Ch

Pagina 53 - Outlet Valve

3 | Preparing the Instrument144 | NGC Chromatography Systems and ChromLab Softwaree. Click Start.4. Run the buffer solution through the colu

Pagina 54 - Ports on the Outlet Valve

Calibrating the NGC InstrumentInstrument Guide | 145Calibrating the NGC Instrument The NGC systems include integrated pressure and temperature sen

Pagina 55 - Outlet Valve LEDs

3 | Preparing the Instrument146 | NGC Chromatography Systems and ChromLab SoftwareCalibrating the Pump Flow RatePrime and purge the system p

Pagina 56 - How the Outlet Valve Works

Calibrating the NGC InstrumentInstrument Guide | 1473. Follow the steps in the Calibration dialog box.4. Verify that Status (at the bottom of the

Pagina 57

Main FeaturesInstrument Guide | 1 3 NGC Discover Pro chromatography system includes All modules on the NGC Discover systemNote: Only the multi-

Pagina 58

3 | Preparing the Instrument148 | NGC Chromatography Systems and ChromLab Software

Pagina 59

Instrument Guide | 149A Maintaining the InstrumentRegular cleaning and maintenance ensures that your NGC™ system continues to perform optimally. T

Pagina 60 - Column Switching Valve LEDs

A | Maintaining the Instrument150 | NGC Chromatography Systems and ChromLab SoftwareCleaning the Outer Surfaces of the NGC InstrumentDuring

Pagina 61 - Instrument Guide

Storing the NGC InstrumentInstrument Guide | 151Storing the NGC InstrumentTip: You can use the System CIP/Storage phase in ChromLab to store the

Pagina 62

A | Maintaining the Instrument152 | NGC Chromatography Systems and ChromLab SoftwareRecommended Maintenance ScheduleTable 14 lists recommen

Pagina 63 - Instrument Guide

Recommended Maintenance ScheduleInstrument Guide | 153YearlySample pump seals and check valveReplace the seals and check valves using the pump hea

Pagina 64

A | Maintaining the Instrument154 | NGC Chromatography Systems and ChromLab SoftwareAs NeededTubing and connectors Replace all tubing and co

Pagina 65

Recommended Cleaning and Storage SolutionsInstrument Guide | 155Recommended Cleaning and Storage SolutionsTable 15 lists the recommended cleaning

Pagina 66 - Detectors

A | Maintaining the Instrument156 | NGC Chromatography Systems and ChromLab SoftwareSystem and Sample PumpsThe NGC system’s gradient pump sy

Pagina 67 - Instrument Guide

System and Sample PumpsInstrument Guide | 157Pump Head Assembly LEGEND1 Outlet check valve housing 2 Pump head small washer3 Pump head seal 4 Inle

Pagina 68 - Single-Wavelength UV Detector

1 | Introduction14 | NGC Chromatography Systems and ChromLab Software Extend the preconfigured systems with additional valves for buffers,

Pagina 69 - Instrument Guide

A | Maintaining the Instrument158 | NGC Chromatography Systems and ChromLab SoftwarePiston Wash Assembly LEGEND1 Fitting 2 Piston base3 Pist

Pagina 70

System and Sample PumpsInstrument Guide | 159Disassembling the PumpsThe system and sample pumps can be disassembled to clean or replace their comp

Pagina 71 - Conductivity Detector

A | Maintaining the Instrument160 | NGC Chromatography Systems and ChromLab Software3. Use the 3/16" hex key that is provided in the fi

Pagina 72 - How They Work

System and Sample PumpsInstrument Guide | 1615. Locate the wash housing O-ring and the pump head small washer. If you do not see them on the separ

Pagina 73 - The Backpressure Regulator

A | Maintaining the Instrument162 | NGC Chromatography Systems and ChromLab Software9. (Optional) Inspect and clean the piston.Note: To rep

Pagina 74

System and Sample PumpsInstrument Guide | 1632. Unscrew each IN check valve housing and pull it down out of the pump head.You should have removed

Pagina 75 - Ports on the pH Detector

A | Maintaining the Instrument164 | NGC Chromatography Systems and ChromLab Software3. Insert two check valve capsules into each IN check va

Pagina 76 - How the pH Detector Works

System and Sample PumpsInstrument Guide | 165To install the pump head check valve assemblies1. To determine the inlet and outlet pump head housing

Pagina 77 - Instrument Guide

A | Maintaining the Instrument166 | NGC Chromatography Systems and ChromLab Software4. Insert an OUT check valve assembly into the open port

Pagina 78 - Mounted

System and Sample PumpsInstrument Guide | 167To replace the pump head seal1. Insert the seal removal end of the small seal tool in the seal cavity

Pagina 79 - Sample Injection Loops

Site RequirementsInstrument Guide | 1 5EnvironmentThe NGC instruments are designed to work within temperatures of 4–40°C.The small footprint makes

Pagina 80 - Fittings

A | Maintaining the Instrument168 | NGC Chromatography Systems and ChromLab Software6. Remove the seal tool and inspect the pump head to ver

Pagina 81 - LEDs and LED Screens

System and Sample PumpsInstrument Guide | 169Removing the Wash HousingTo remove the wash housing1. Complete the steps in the section Removing the

Pagina 82 - Air Sensors

A | Maintaining the Instrument170 | NGC Chromatography Systems and ChromLab SoftwareRemoving the PistonCaution: Use extreme caution when re

Pagina 83 - Instrument Guide

System and Sample PumpsInstrument Guide | 1715. Gently grasp the piston by its metal base and carefully lift the piston straight up to dislodge it

Pagina 84 - Air Sensor LEDs

A | Maintaining the Instrument172 | NGC Chromatography Systems and ChromLab Software6. Rinse the piston and examine it for damage.7. Store t

Pagina 85 - Temperature Sensor

System and Sample PumpsInstrument Guide | 173To replace the wash housing check valves1. Identify the left and right wash housings. The wash syste

Pagina 86 - NGC Touch Screen

A | Maintaining the Instrument174 | NGC Chromatography Systems and ChromLab SoftwareTo identify the left and right housings after you remove

Pagina 87 - Buffer Tray

System and Sample PumpsInstrument Guide | 175Reassembling the PumpsThis section explains how to Insert the piston Install the wash housing Inst

Pagina 88 - Product Configurations

A | Maintaining the Instrument176 | NGC Chromatography Systems and ChromLab SoftwareInstalling the Wash HousingBefore installing the wash ho

Pagina 89 - Discover series

System and Sample PumpsInstrument Guide | 1774. Position the left wash housing with its arrow pointing up and carefully slide it into place over t

Pagina 90

1 | Introduction16 | NGC Chromatography Systems and ChromLab SoftwareSafety RequirementsThis section explains the different safety alerts in

Pagina 91 - 3 Preparing the Instrument

A | Maintaining the Instrument178 | NGC Chromatography Systems and ChromLab SoftwareInstalling the Pump HeadsNote: When installing new pump

Pagina 92 - Module Review

System and Sample PumpsInstrument Guide | 1796. Identify the pump head assemblies. To identify the left and right pump heads, position both pump h

Pagina 93 - Instrument Guide

A | Maintaining the Instrument180 | NGC Chromatography Systems and ChromLab Software8. Push on the center of the pump head to seat it firmly

Pagina 94 - NGC Systems Accessories Kit

System and Sample PumpsInstrument Guide | 181Replacing the Priming Manifold Check ValveThe priming manifold check valve has one check valve capsul

Pagina 95 - NGC Discover Accessories Kit

A | Maintaining the Instrument182 | NGC Chromatography Systems and ChromLab Software2. Using a pair of tweezers, remove the check valve caps

Pagina 96 - Waste Bottles

System and Sample PumpsInstrument Guide | 1835. Using the tweezers, insert the new check valve capsule into the priming manifold outlet port.Note:

Pagina 97 - Instrument Guide

A | Maintaining the Instrument184 | NGC Chromatography Systems and ChromLab Software6. Carefully insert the OUT check valve housing in the p

Pagina 98

System and Sample PumpsInstrument Guide | 185Conditioning New SealsNew seals should be conditioned before use. To condition the seals, run the pum

Pagina 99 - Preparing the Detectors

A | Maintaining the Instrument186 | NGC Chromatography Systems and ChromLab SoftwareMixerReplacing the Mixer ComponentsThe NGC system’s mixi

Pagina 100 - Changing the Mixer Barrel

MixerInstrument Guide | 187

Pagina 101 - F10 Pump F100 Pump

Finding Out MoreInstrument Guide | 1 7Finding Out MoreAfter you install NGC documentation from the NGC Chromatography Systems Documentation DVD, y

Pagina 102

A | Maintaining the Instrument188 | NGC Chromatography Systems and ChromLab SoftwareDisassembling the MixerUse this procedure to replace any

Pagina 103 - (air sensor to pump inlet)

MixerInstrument Guide | 1893. Twist the mixer assembly clockwise approximately 1/8 turn and lift it upward to remove it from the socket.4. Tilt th

Pagina 104

A | Maintaining the Instrument190 | NGC Chromatography Systems and ChromLab Software6. Separate the mixer top from the mixer base and barrel

Pagina 105 - To activate the air sensors

MixerInstrument Guide | 191To change the O-rings1. Locate the O-ring in the groove on the mixer base and the mixer barrel extension (if used).2. W

Pagina 106

A | Maintaining the Instrument192 | NGC Chromatography Systems and ChromLab SoftwareTo replace or exchange the mixer barrelTip: If you are

Pagina 107 - Instrument Guide

MixerInstrument Guide | 193To clean the mixer body1. Wipe the inside of the mixer body with a lint-free towel.2. Soak and sonicate the following c

Pagina 108

A | Maintaining the Instrument194 | NGC Chromatography Systems and ChromLab SoftwareReassembling the MixerTip: To reinsert the magnetic st

Pagina 109 - Instrument Guide

MixerInstrument Guide | 1953. Using the hex key, firmly tighten the hex bolts. Alternate by tightening opposite bolts (tighten Bolt 1, then Bolt 3

Pagina 110

A | Maintaining the Instrument196 | NGC Chromatography Systems and ChromLab SoftwareThe DetectorsReplacing the UV Flow CellThe NGC systems s

Pagina 111 - Instrument Guide

The DetectorsInstrument Guide | 1974. Firmly grasp the bolts on the flow cell and lift the flow cell from its socket.Multi-Wavelength UV/Vis Flow

Pagina 113 - Instrument Guide

1 | Introduction18 | NGC Chromatography Systems and ChromLab Software

Pagina 114

A | Maintaining the Instrument198 | NGC Chromatography Systems and ChromLab Software5. Verify that the square gasket is removed with the flo

Pagina 115 - Positive

The DetectorsInstrument Guide | 1997. Locate the flow cell. The flow cell should include Two bolts One square rubber gasket8. Place the flow cel

Pagina 116 - Starting the NGC Instrument

A | Maintaining the Instrument200 | NGC Chromatography Systems and ChromLab Software9. Tighten the bolts. If necessary, use the screwdriver.

Pagina 117 - Instrument Guide

The DetectorsInstrument Guide | 201Replacing the Single-Wavelength UV Detector LEDThis section explains how to replace the LED on the single-wavel

Pagina 118

A | Maintaining the Instrument202 | NGC Chromatography Systems and ChromLab Software

Pagina 119 - To verify the plumbing path

The DetectorsInstrument Guide | 203Tip: The door is tethered to the casing. It will hang to the side while you replace the lamps.

Pagina 120

A | Maintaining the Instrument204 | NGC Chromatography Systems and ChromLab SoftwareYou see the LED module with three thumbscrews. The LED i

Pagina 121 - To prime the system pumps

The DetectorsInstrument Guide | 2056. Loosen the left thumbscrews, alternating one turn on the top left screw then one turn on the bottom left scr

Pagina 122 - To purge the system pumps

A | Maintaining the Instrument206 | NGC Chromatography Systems and ChromLab SoftwareThe following image shows the LED housing without the LE

Pagina 123 - Instrument Guide

The DetectorsInstrument Guide | 20710. Align the thumbscrews of the replacement LED with the holes on the housing and carefully insert the LED int

Pagina 124 - Priming the NGC Scout System

Instrument Guide | 1 92 The NGC InstrumentThe NGC™ instrument ships preassembled with the components necessary to perform gradient separations. Th

Pagina 125 - Instrument Guide

A | Maintaining the Instrument208 | NGC Chromatography Systems and ChromLab SoftwareResetting the Lamp Time DetailsThe system tracks the lam

Pagina 126 - Purging the NGC Scout System

The DetectorsInstrument Guide | 209Replacing the Multi-Wavelength UV/Vis Detector Lamps This section explains how to replace the deuterium lamp or

Pagina 127 - Instrument Guide

A | Maintaining the Instrument210 | NGC Chromatography Systems and ChromLab SoftwareThe deuterium and tungsten lamps are both accessible.Tip

Pagina 128 - Priming NGC Discover Systems

The DetectorsInstrument Guide | 2116. Remove the lamp:WARNING! Use caution when removing the lamps. The lamps might be hot to the touch after th

Pagina 129 - Instrument Guide

A | Maintaining the Instrument212 | NGC Chromatography Systems and ChromLab Softwarec. With a 2.5 mm hex key, remove the screws on the top a

Pagina 130

The DetectorsInstrument Guide | 21311. Insert the cable plug and tighten the collar.Tip: The deuterium lamp plug has seven connector pins, the t

Pagina 131 - Instrument Guide

A | Maintaining the Instrument214 | NGC Chromatography Systems and ChromLab SoftwareReplacing the Conductivity MonitorThe images in this sec

Pagina 132 - To prime the sample pump

The DetectorsInstrument Guide | 2154. Firmly grasp the detector and pull it toward you to detach it.5. Dispose of the detector in an appropriate w

Pagina 133 - Instrument Guide

A | Maintaining the Instrument216 | NGC Chromatography Systems and ChromLab Software8. Gently push the detector onto the connector pins on t

Pagina 134

pH ProbeInstrument Guide | 217pH ProbeStoring the pH ElectrodeWhen not in use, store the pH electrode in storage solution. See Table 15 on page

Pagina 135 - Instrument Guide

2 | The NGC Instrument20 | NGC Chromatography Systems and ChromLab SoftwareThe NGC Instrument IllustratedThe illustrations in this section d

Pagina 136

A | Maintaining the Instrument218 | NGC Chromatography Systems and ChromLab SoftwareTo remove lipophilic organic contaminants Immerse the e

Pagina 137 - Instrument Guide

pH ProbeInstrument Guide | 2193. Loosen the black collar on the pH probe.4. To remove the pH probe, lift the probe out of the pH flow cell.Dispose

Pagina 138 - To purge the sample pump

A | Maintaining the Instrument220 | NGC Chromatography Systems and ChromLab SoftwareOther ComponentsAttaching an Expansion Tier to the NGC I

Pagina 139 - Instrument Guide

Other ComponentsInstrument Guide | 221Tip: The buffer tray can remain inside the top.4. Locate the package that contains the expansion tier and

Pagina 140

A | Maintaining the Instrument222 | NGC Chromatography Systems and ChromLab Software7. Locate the connector cable that ships with the expans

Pagina 141 - Instrument Guide

Other ComponentsInstrument Guide | 223For example:

Pagina 142 - Preparing the Columns

A | Maintaining the Instrument224 | NGC Chromatography Systems and ChromLab Software9. Firmly press the clips inward and insert the appropri

Pagina 143 - Instrument Guide

Other ComponentsInstrument Guide | 22510. Insert the other connector into the lower tier’s connector port.11. Place the top, with the buffer tray,

Pagina 144 -  UV monitor

A | Maintaining the Instrument226 | NGC Chromatography Systems and ChromLab Software13. Perform the following tasks:Task: Refer to the secti

Pagina 145 - Washing the Columns

Other ComponentsInstrument Guide | 227Replacing or Repositioning Modules on the NGC InstrumentsWARNING! Disconnect power to the NGC instrument b

Pagina 146

This page is intentionally blank.The NGC Instrument IllustratedInstrument Guide | 2 1

Pagina 147 - Calibrating pH

A | Maintaining the Instrument228 | NGC Chromatography Systems and ChromLab SoftwareThe following image shows an empty bay.5. Store the modu

Pagina 148 - To zero the pressure

Other ComponentsInstrument Guide | 2298. The frame of the bay has tracks on the top and bottom of the right side. The module has guide rails on th

Pagina 149 - Instrument Guide

A | Maintaining the Instrument230 | NGC Chromatography Systems and ChromLab Software13. Perform the following tasks:Task: Refer to the secti

Pagina 150

Other ComponentsInstrument Guide | 231Converting Bays to Fit ModulesSome modules fit into single-wide bays while others require double-wide bays (

Pagina 151 - A Maintaining the Instrument

A | Maintaining the Instrument232 | NGC Chromatography Systems and ChromLab SoftwareTo convert a single-wide bay to a double-wide bay1. Comp

Pagina 152

Other ComponentsInstrument Guide | 233To convert a double-wide bay to a single-wide bay1. Locate a divider.2. Complete steps 1–5 in the section To

Pagina 153 - Storing the NGC Instrument

A | Maintaining the Instrument234 | NGC Chromatography Systems and ChromLab SoftwareCutting Replacement TubingTip: Use the tubing cutter p

Pagina 154 - Interval Component Action

Other ComponentsInstrument Guide | 2354. Slide the fitting nut, lock ring, and the ferrule, in that order, onto the tubing.Note: The stainless st

Pagina 155

A | Maintaining the Instrument236 | NGC Chromatography Systems and ChromLab SoftwareInstalling the Backpressure RegulatorThe NGC systems shi

Pagina 156 - As Needed

Other ComponentsInstrument Guide | 237To remove the backpressure regulator1. Disconnect the tubing from the top and bottom of the installed backpr

Pagina 157 - Component Solution to use

2 | The NGC Instrument22 | NGC Chromatography Systems and ChromLab SoftwareFront View

Pagina 158 - System and Sample Pumps

A | Maintaining the Instrument238 | NGC Chromatography Systems and ChromLab SoftwareRepositioning the Touch ScreenThis procedure explains ho

Pagina 159 - Pump Head Assembly

Other ComponentsInstrument Guide | 2394. Remove the side mount cover from the side of the instrument on which you plan to attach the touch screen.

Pagina 160 - Piston Wash Assembly

A | Maintaining the Instrument240 | NGC Chromatography Systems and ChromLab Software6. Remove the display cable from the side and rear cable

Pagina 161 - Disassembling the Pumps

Other ComponentsInstrument Guide | 2418. Lift the touch screen display off the mounting post.9. Place the touch screen monitor face down on a clea

Pagina 162

A | Maintaining the Instrument242 | NGC Chromatography Systems and ChromLab Software11. Using the 7/64 hex key provided in the accessories k

Pagina 163 - Instrument Guide

Other ComponentsInstrument Guide | 24314. Turn the pivot arm assembly around 180° so that the mounting assembly is now in your left hand and inser

Pagina 164 - To clean the pump heads

A | Maintaining the Instrument244 | NGC Chromatography Systems and ChromLab Software20. Route the display cable through the side and rear ca

Pagina 165 - Instrument Guide

Instrument Guide | 245B Troubleshooting the InstrumentThis appendix lists potential problems and suggested solutions for the NGC™ chromatography s

Pagina 166

B | Troubleshooting the Instrument246 | NGC Chromatography Systems and ChromLab SoftwareThe pumps do not deliver the correct flow rate (cont

Pagina 167 - Instrument Guide

System and Sample PumpsInstrument Guide | 247The pumps do not deliver the correct flow rate (continued).Air bubbles might be trapped in the pump h

Pagina 168

The NGC Instrument IllustratedInstrument Guide | 2 3LEGEND1 Touch screen monitor 2 Sample pump3 Sample inlet valve 4 Buffer tray5 Empty bay, cover

Pagina 169 - To replace the pump head seal

B | Troubleshooting the Instrument248 | NGC Chromatography Systems and ChromLab SoftwareThe pumps do not deliver the correct flow rate (cont

Pagina 170

System and Sample PumpsInstrument Guide | 249Liquid continuously leaks from the pump head washout drain trough (continued).The check valves might

Pagina 171 - Removing the Wash Housing

B | Troubleshooting the Instrument250 | NGC Chromatography Systems and ChromLab SoftwareThe backpressure drops to zero or is much lower than

Pagina 172 - Removing the Piston

System and Sample PumpsInstrument Guide | 251Backpressure values are higher or lower than expected. The flow rate might have changed. In ChromLab,

Pagina 173 - Instrument Guide

B | Troubleshooting the Instrument252 | NGC Chromatography Systems and ChromLab SoftwareBackpressure values are higher or lower than expecte

Pagina 174

UV or UV/Vis DetectorInstrument Guide | 253UV or UV/Vis DetectorSymptom Possible Cause Possible SolutionThe UV baseline is unstable or noisy.The U

Pagina 175 - Instrument Guide

B | Troubleshooting the Instrument254 | NGC Chromatography Systems and ChromLab SoftwareThe UV baseline is unstable or noisy (continued).The

Pagina 176

UV or UV/Vis DetectorInstrument Guide | 255Baseline noise continues when pumps are turned off.External environmental influences can cause noise sp

Pagina 177 - Reassembling the Pumps

B | Troubleshooting the Instrument256 | NGC Chromatography Systems and ChromLab SoftwareThe baseline is drifting. The eluant is nonhomogeneo

Pagina 178 - Installing the Wash Housing

UV or UV/Vis DetectorInstrument Guide | 257Sample trace values are lower than expected.The lamps might be nearing the end of their usable life.Che

Pagina 179 - Instrument Guide

2 | The NGC Instrument24 | NGC Chromatography Systems and ChromLab SoftwareBack ViewLEGEND1 Cable connector port/hand grip 2 Cable connector

Pagina 180 - Installing the Pump Heads

B | Troubleshooting the Instrument258 | NGC Chromatography Systems and ChromLab SoftwareConductivity MonitorSymptom Possible Cause Possible

Pagina 181 - Instrument Guide

pH ProbeInstrument Guide | 259pH ProbeSymptom Possible Cause Possible SolutionThe pH readings are higher or lower than expected.The pH probe might

Pagina 182

B | Troubleshooting the Instrument260 | NGC Chromatography Systems and ChromLab SoftwareBuffer Blending ValveColumn Switching ValveSymptom P

Pagina 183 - Instrument Guide

External DevicesInstrument Guide | 261External DevicesSymptom Possible Cause Possible SolutionUnable to zero an external device connected to SIM 1

Pagina 184

B | Troubleshooting the Instrument262 | NGC Chromatography Systems and ChromLab Software

Pagina 185 - Instrument Guide

Instrument Guide | 263C Plumbing the SystemWith the exception of the inlet tubing, waste lines, and column, the NGC™ systems ship preassembled and

Pagina 186

C | Plumbing the System264 | NGC Chromatography Systems and ChromLab SoftwarePlumbing NGC SystemsSeveral plumbing tasks are required for all

Pagina 187 - Conditioning New Seals

Plumbing NGC SystemsInstrument Guide | 265Plumbing All NGC SystemsTable 19 lists the tasks for plumbing all NGC systems. The tasks are explained

Pagina 188

C | Plumbing the System266 | NGC Chromatography Systems and ChromLab SoftwareSystem Pumps Wash FlowSample Pump Wash Flow

Pagina 189 - Instrument Guide

Plumbing NGC SystemsInstrument Guide | 267To plumb and prime the system pump wash lines1. Mount a tube holder to the right or left side of the ins

Pagina 190 - Disassembling the Mixer

The NGC Instrument IllustratedInstrument Guide | 2 5Right Side ViewLEGEND1 Drain hole 2 Peripheral port 3 Peripheral ports 4 USB ports5 Soft power

Pagina 191 - Instrument Guide

C | Plumbing the System268 | NGC Chromatography Systems and ChromLab Software8. Attach a 30 ml syringe to the 1/4-28 female to male luer ada

Pagina 192

Plumbing NGC SystemsInstrument Guide | 269Plumbing the Sample Inject ValveTo plumb the sample inject valve1. Connect waste tubing to waste ports W

Pagina 193 - To change the O-rings

C | Plumbing the System270 | NGC Chromatography Systems and ChromLab SoftwareAttaching the Backpressure RegulatorIf your column requires a h

Pagina 194

Plumbing NGC SystemsInstrument Guide | 271Plumbing the NGC Quest SystemTable 20 lists the tasks for plumbing the NGC Quest™ system. The tasks are

Pagina 195 - To clean the mixer body

C | Plumbing the System272 | NGC Chromatography Systems and ChromLab SoftwareNGC Quest Point-to-Plumb Connections

Pagina 196 - Reassembling the Mixer

Plumbing NGC SystemsInstrument Guide | 273To plumb the system pumps1. Connect inlet tubing to the inlet manifold of system pump A.2. Connect inlet

Pagina 197 - Instrument Guide

C | Plumbing the System274 | NGC Chromatography Systems and ChromLab SoftwarePlumbing the NGC Scout SystemTable 21 lists the tasks for plum

Pagina 198 - The Detectors

Plumbing NGC SystemsInstrument Guide | 275NGC Scout Point-to-Plumb Connections

Pagina 199 - Instrument Guide

C | Plumbing the System276 | NGC Chromatography Systems and ChromLab SoftwareTo plumb the buffer blending valve Connect the inlet tubing to

Pagina 200

Plumbing NGC SystemsInstrument Guide | 2773. Immerse the open end of the waste tubing in the waste container.4. Connect Tube #2 from the 40 psi ba

Pagina 201 - Instrument Guide

2 | The NGC Instrument26 | NGC Chromatography Systems and ChromLab SoftwareLeft Side ViewLEGEND1 Buffer tray 2 Drain hole3 Touch screen cabl

Pagina 202

C | Plumbing the System278 | NGC Chromatography Systems and ChromLab SoftwarePlumbing the NGC Discover SystemTable 22 lists the tasks for p

Pagina 203 - Instrument Guide

Plumbing NGC SystemsInstrument Guide | 279NGC Discover Point-to-Plumb Connections

Pagina 204

C | Plumbing the System280 | NGC Chromatography Systems and ChromLab SoftwareTo plumb the buffer blending valve Connect the inlet tubing to

Pagina 205 - Instrument Guide

Plumbing NGC SystemsInstrument Guide | 281To plumb the buffer inlet valves1. Connect tubing from Outlet A on the buffer blending valve to port 8/B

Pagina 206

C | Plumbing the System282 | NGC Chromatography Systems and ChromLab SoftwareTo plumb the pH valve1. Connect the pump head wash priming tubi

Pagina 207 - Instrument Guide

Plumbing NGC SystemsInstrument Guide | 283Plumbing the NGC Discover Pro SystemTable 23 lists the tasks for plumbing the NGC Discover Pro system.

Pagina 208

C | Plumbing the System284 | NGC Chromatography Systems and ChromLab SoftwareNGC Discover Pro Point-to-Plumb Connections

Pagina 209 - Front of LED Board

Plumbing NGC SystemsInstrument Guide | 285Plumbing Sample Inlet ValvesEach sample inlet valve module includes all necessary tubing. The inlet tubi

Pagina 210 - To reset lamp time details

C | Plumbing the System286 | NGC Chromatography Systems and ChromLab Software2. If your system has two sample inlet valves:a. Connect tubing

Pagina 211 - Instrument Guide

Plumbing NGC SystemsInstrument Guide | 287b. (Optional) Do one of the following: Connect Tube #4 from port 1 to the Common inlet port on the BioF

Pagina 212

The NGC Instrument IllustratedInstrument Guide | 2 7Liquid Flow Path of the NGC Discover Pro SystemAt startup, ChromLab displays a default fluidic

Pagina 213 - 6. Remove the lamp:

C | Plumbing the System288 | NGC Chromatography Systems and ChromLab Software

Pagina 214

Instrument Guide | 289D Connecting the C-96 Autosampler to NGC SystemsUsing the C-96 Autosampler with NGC SystemsThe C-96 autosampler enhances NGC

Pagina 215 - Instrument Guide

D | Connecting the C-96 Autosampler to NGC Systems290 | NGC Chromatography Systems and ChromLab Software5. Locate the two cables that conn

Pagina 216

Using the C-96 Autosampler with NGC SystemsInstrument Guide | 2918. On the back of the autosampler:a. Insert the 15-pin cable into the left (P1 RE

Pagina 217 - Instrument Guide

D | Connecting the C-96 Autosampler to NGC Systems292 | NGC Chromatography Systems and ChromLab SoftwarePlumbing and Priming the C-96 Auto

Pagina 218

Using the C-96 Autosampler with NGC SystemsInstrument Guide | 293To prime and purge the C-96 autosamplerNote: Before priming the autosampler, ens

Pagina 219 - Cleaning the pH Electrode

D | Connecting the C-96 Autosampler to NGC Systems294 | NGC Chromatography Systems and ChromLab Software7. In the fluidic scheme, touch or

Pagina 220 - Replacing the pH Probe

Using the C-96 Autosampler with NGC SystemsInstrument Guide | 295Setting Up the C-96 Autosampler for Use in ChromLab MethodsThis section briefly e

Pagina 221 - Instrument Guide

D | Connecting the C-96 Autosampler to NGC Systems296 | NGC Chromatography Systems and ChromLab Software

Pagina 222 - Other Components

Instrument Guide | 297E Regulatory InformationThe NGC™ instrument has been tested and found to be in compliance with all applicable requirements o

Pagina 223 - Instrument Guide

NGC™ Chromatography Systems and ChromLab™ SoftwareInstrument GuideVersion 3.0

Pagina 224 - Previously attached

2 | The NGC Instrument28 | NGC Chromatography Systems and ChromLab SoftwarePumpsThe NGC systems can have up to three high-precision pumps: t

Pagina 225 - For example:

E | Regulatory Information298 | NGC Chromatography Systems and ChromLab SoftwareSample Loop PSI Ratings for U.S. and CanadaThe following pre

Pagina 226

FCC Warning and NotesInstrument Guide | 299FCC Warning and NotesThe NGC instrument complies with part 15 of the FCC Rules. Operation is subject to

Pagina 227 - Instrument Guide

E | Regulatory Information300 | NGC Chromatography Systems and ChromLab Software

Pagina 228 - Task: Refer to the section:

Instrument Guide | 301F NGC Chromatography Systems Catalog NumbersTable 25. NGC™ chromatography systems catalog numbersSystem Catalog numberNGC

Pagina 229 - Instruments

F | NGC Chromatography Systems Catalog Numbers302 | NGC Chromatography Systems and ChromLab Software

Pagina 231 - Top of Module

Life ScienceGroup14-0273 0114 Sig 121310026253 Rev C US/EGBio-Rad Laboratories, Inc.Web site www.bio-rad.com USA 800 424 6723 Australia 61

Pagina 232

PumpsInstrument Guide | 2 9System PumpsThe system pumps perform isocratic or gradient elution at the specified flow rates.LEGEND1 Pump flow status

Pagina 233 - Instrument Guide

2 | The NGC Instrument30 | NGC Chromatography Systems and ChromLab SoftwareDetailed AttributesThe NGC systems include two system pumps: pump

Pagina 234

PumpsInstrument Guide | 3 1Each system pump module has an emergency Pause button and a Purge button. Pressing the emergency Pause button on eithe

Pagina 235 - Instrument Guide

2 | The NGC Instrument32 | NGC Chromatography Systems and ChromLab SoftwarePiston Wash SystemThe pumps on the NGC system automatically rinse

Pagina 236 - Cutting Replacement Tubing

PumpsInstrument Guide | 3 3Sample PumpThe sample pump is used to automatically load large sample volumes onto a column or to automatically fill la

Pagina 237 - Instrument Guide

2 | The NGC Instrument34 | NGC Chromatography Systems and ChromLab SoftwareDetailed Attributes Flow rate delivery: 0.01–100 ml/min Pressur

Pagina 238

PumpsInstrument Guide | 3 5Tip: You can pause the system through ChromLab using either the touch screen or the ChromLab computer. You can also s

Pagina 239 - Instrument Guide

2 | The NGC Instrument36 | NGC Chromatography Systems and ChromLab SoftwareValvesAll NGC systems include a sample inject valve. Your system

Pagina 240

ValvesInstrument Guide | 3 7Sample Inject ValveThe sample inject valve enables the system to load a specific, predetermined volume of sample onto

Pagina 241 - Instrument Guide

Bio-Rad Technical Support DepartmentThe Bio-Rad Technical Support department in the U.S. is open Monday through Friday, 5:00 AM to 5:00 PM, Pacific Ti

Pagina 242

2 | The NGC Instrument38 | NGC Chromatography Systems and ChromLab SoftwarePorts on the Sample Inject Valve Port FunctionInject Manually loa

Pagina 243 - Instrument Guide

ValvesInstrument Guide | 3 9Sample Inject Valve LEDsThe sample inject valve module is fitted with LEDs. Solid green LEDs indicate the flow is fro

Pagina 244

2 | The NGC Instrument40 | NGC Chromatography Systems and ChromLab SoftwareValve Position ExplanationManual Load Loop/System Pump to Column

Pagina 245 - Instrument Guide

ValvesInstrument Guide | 4 1Sample Pump Direct Inject/System Pump Waste Directs system pump flow to waste (W2) Directs sample pump flow directly

Pagina 246

2 | The NGC Instrument42 | NGC Chromatography Systems and ChromLab SoftwareBuffer Blending Valve The buffer blending valve is used for autom

Pagina 247 - Instrument

ValvesInstrument Guide | 4 3Detailed Attributes Operating pressure up to 15 psi

Pagina 248

2 | The NGC Instrument44 | NGC Chromatography Systems and ChromLab SoftwarePorts on the Buffer Blending Valve Buffer Blending Valve LEDsThe

Pagina 249 - Instrument Guide

ValvesInstrument Guide | 4 5How the Buffer Blending Valve WorksThe buffer blending valve operates in two modes: buffer blending and gradient.Buffe

Pagina 250 -  Replace the check valves

2 | The NGC Instrument46 | NGC Chromatography Systems and ChromLab SoftwareInlet ValvesThe inlet valves automate the use of multiple buffers

Pagina 251 - Instrument Guide

ValvesInstrument Guide | 4 7Inlet ValveDetailed Attributes Maximum operating pressure: 500 psi  Optional air sensors indicate the end of buffer

Pagina 252 -  Replace check valves

Instrument Guide | iiiTable of ContentsChapter 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11The N

Pagina 253 - Instrument Guide

2 | The NGC Instrument48 | NGC Chromatography Systems and ChromLab SoftwarePorts on the Inlet ValvesBuffer Inlet ValveSample Inlet ValvePort

Pagina 254

ValvesInstrument Guide | 4 9Inlet Valve LEDsThe inlet valve modules are fitted with LEDs. Solid green LEDs indicate the flow is from the system p

Pagina 255 - UV or UV/Vis Detector

2 | The NGC Instrument50 | NGC Chromatography Systems and ChromLab SoftwareHow the Inlet Valves WorkThe function of an inlet valve is determ

Pagina 256 -  See Replacing the Multi

ValvesInstrument Guide | 5 1Outlet ValveThe outlet valve enables the system to direct the flow to the BioFrac™ fraction collector, to waste, or to

Pagina 257 - Instrument Guide

2 | The NGC Instrument52 | NGC Chromatography Systems and ChromLab SoftwareDetailed Attributes Maximum operating pressure is 500 psiPorts o

Pagina 258 -  Recalibrate both system

ValvesInstrument Guide | 5 3Outlet Valve LEDsThe outlet valve modules are fitted with LEDs. Solid green LEDs indicate the flow is from the system

Pagina 259 - Instrument Guide

2 | The NGC Instrument54 | NGC Chromatography Systems and ChromLab SoftwareHow the Outlet Valve WorksThe NGC system supports the use of up t

Pagina 260 - Conductivity Monitor

ValvesInstrument Guide | 5 5Column Switching ValveThe NGC system supports up to three column switching valves. Each column switching valve is used

Pagina 261 - Instrument Guide

2 | The NGC Instrument56 | NGC Chromatography Systems and ChromLab SoftwareDetailed Attributes Maximum operating pressure: 3,650 psi  Dual

Pagina 262 - Column Switching Valve

ValvesInstrument Guide | 5 7Ports on the Column Switching Valve Port FunctionTop 1–5 Ports to connect to the top of columnsBottom 1–5 Ports

Pagina 263 - External Devices

iv | NGC Chromatography Systems and ChromLab SoftwareTable of ContentsHow the Sample Pump Works . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 264

2 | The NGC Instrument58 | NGC Chromatography Systems and ChromLab SoftwareColumn Switching Valve LEDsThe column switching valve modules are

Pagina 265 - C Plumbing the System

ValvesInstrument Guide | 5 9The section Flow Paths of the Column Switching Valve explains the specific flow paths.The column switching valve has i

Pagina 266 - Plumbing NGC Systems

2 | The NGC Instrument60 | NGC Chromatography Systems and ChromLab Software Select Bypass to bypass the flow past all ports. Select Revers

Pagina 267 - Plumbing All NGC Systems

MixerInstrument Guide | 6 1Mixer The mixer module consists of a mixer assembly and an integrated system pressure sensor. This module homogenizes t

Pagina 268 - Sample Pump Wash Flow

2 | The NGC Instrument62 | NGC Chromatography Systems and ChromLab SoftwareDuring operation, the system pressure sensor continuously measure

Pagina 269 - Instrument Guide

MixerInstrument Guide | 6 3Detailed AttributesThe mixer is composed of a base motor, a stirring plate, and a hollow mixer barrel in which the buff

Pagina 270 - To plumb the sample pump

2 | The NGC Instrument64 | NGC Chromatography Systems and ChromLab SoftwareMixer LEDsThe mixer module is fitted with LEDs. A left green LED

Pagina 271 - Instrument Guide

DetectorsInstrument Guide | 6 5Single-Wavelength UV Detector and Conductivity MonitorMulti-Wavelength UV Detector and Conductivity Monitor

Pagina 272 - C

2 | The NGC Instrument66 | NGC Chromatography Systems and ChromLab SoftwareSingle-Wavelength UV DetectorDetailed Attributes Two integrated

Pagina 273 - Plumbing the NGC Quest System

DetectorsInstrument Guide | 6 7 Monitors UV absorbance one wavelength at a time 255 nm  280 nm (default)Tip: Use ChromLab to change the absor

Pagina 274

Instrument Guide | vTable of ContentsMulti-Wavelength UV/Vis Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68Conductiv

Pagina 275 - To plumb the system pumps

2 | The NGC Instrument68 | NGC Chromatography Systems and ChromLab SoftwareMulti-Wavelength UV/Vis DetectorDetailed Attributes Contains one

Pagina 276 - Plumbing the NGC Scout System

DetectorsInstrument Guide | 6 9 Monitors up to four wavelengths in the UV/Vis range of 190–800 nmTip: Use ChromLab to change the absorbance wav

Pagina 277 - Instrument Guide

2 | The NGC Instrument70 | NGC Chromatography Systems and ChromLab SoftwareHow They WorkUV DetectorsChromLab recognizes the connected optic

Pagina 278 - To plumb the pH valve

DetectorsInstrument Guide | 7 1The Backpressure RegulatorA backpressure regulator can be added to the flow path to generate backpressure and preve

Pagina 279 - Instrument Guide

2 | The NGC Instrument72 | NGC Chromatography Systems and ChromLab SoftwarepH DetectorThe pH detector includes an integrated flow cell in wh

Pagina 280

DetectorsInstrument Guide | 7 3Detailed Attributes Flow cell volume: 90 μl (200 μl including internal valve passageways) Electrode maximum opera

Pagina 281 - Instrument Guide

2 | The NGC Instrument74 | NGC Chromatography Systems and ChromLab SoftwareHow the pH Detector WorksThe integrated pH electrode enables inli

Pagina 282 - Tubing color Port Solution

DetectorsInstrument Guide | 7 5Changing the Flow Paths of the pH ValveDuring a manual run, you can change the flow path of the pH valve by accessi

Pagina 283

2 | The NGC Instrument76 | NGC Chromatography Systems and ChromLab SoftwareTubing, Loops, Columns, and FittingsThis section lists the access

Pagina 284

Tubing, Loops, Columns, and FittingsInstrument Guide | 7 7Sample Injection LoopsColumnsThe NGC chromatography systems support columns for various

Pagina 285 - Instrument Guide

vi | NGC Chromatography Systems and ChromLab SoftwareTable of ContentsConnecting the Sample Injection Loops . . . . . . . . . . . . . . . . . .

Pagina 286

2 | The NGC Instrument78 | NGC Chromatography Systems and ChromLab SoftwareFittingsThe NGC chromatography systems ship with 10/32 fittings f

Pagina 287 - Plumbing Sample Inlet Valves

LEDs and LED ScreensInstrument Guide | 7 9LEDs and LED ScreensThe modules on the NGC system are fitted with light emitting diodes (LEDs) that are

Pagina 288 - Plumbing Outlet Valves

2 | The NGC Instrument80 | NGC Chromatography Systems and ChromLab SoftwareSensorsAir SensorsThe NGC chromatography systems can be fitted wi

Pagina 289 - Instrument Guide

SensorsInstrument Guide | 8 1The air sensors are inserted into ports on the air sensor module. ChromLab detects the presence or absence of air sen

Pagina 290

2 | The NGC Instrument82 | NGC Chromatography Systems and ChromLab SoftwareAir Sensor LEDsThe air sensor module is fitted with LEDs. Green

Pagina 291 - D Connecting the C-96

SensorsInstrument Guide | 8 3Pressure SensorsThe NGC systems can have up to four integrated pressure sensors: System pressure sensor — located in

Pagina 292

2 | The NGC Instrument84 | NGC Chromatography Systems and ChromLab SoftwareNGC Touch ScreenThe NGC system’s touch screen is the main interfa

Pagina 293 - Instrument Guide

Buffer TrayInstrument Guide | 8 5Buffer TrayThe buffer tray on the top of the instrument can accommodate up to six 1 L bottles.Note: Ensure that

Pagina 294 - To prepare the tubing

2 | The NGC Instrument86 | NGC Chromatography Systems and ChromLab SoftwareProduct ConfigurationsThe NGC chromatography system is available

Pagina 295 - Sample Inject

Product ConfigurationsInstrument Guide | 8 7Fourth expansion tierSample inlet valve 1Outlet valve 1*Only the multi-wavelength UV/Vis detector i

Pagina 296

Instrument Guide | viiTable of ContentsAppendix A Maintaining the Instrument . . . . . . . . . . . . . . . . . . . . 149Cleaning the Outer Surfa

Pagina 297 - Instrument Guide

2 | The NGC Instrument88 | NGC Chromatography Systems and ChromLab Software

Pagina 298

Instrument Guide | 8 93 Preparing the InstrumentThe NGC™ instruments ship preassembled with the modules necessary to perform chromatographic separ

Pagina 299 - E Regulatory Information

3 | Preparing the Instrument90 | NGC Chromatography Systems and ChromLab SoftwareModule ReviewThe NGC systems are shipped in one of three st

Pagina 300 - DynaLoop 90 ml F100 270 psi

Module ReviewInstrument Guide | 9 1NGC Discover™ Chromatography SystemNGC Discover Pro Chromatography System

Pagina 301 - FCC Warning and Notes

3 | Preparing the Instrument92 | NGC Chromatography Systems and ChromLab SoftwareNGC Systems Accessories KitEach NGC system ships with an ac

Pagina 302

NGC Systems Accessories KitInstrument Guide | 9 3NGC Scout Accessories KitThe kit for the NGC Scout system includes all the accessories in the NGC

Pagina 303 - Systems Catalog Numbers

3 | Preparing the Instrument94 | NGC Chromatography Systems and ChromLab SoftwarePreparing the Buffer and Waste BottlesAfter you plumb the i

Pagina 304

Connecting the Sample Injection LoopsInstrument Guide | 9 5Connecting the Sample Injection LoopsThe NGC chromatography systems support a variety o

Pagina 305

3 | Preparing the Instrument96 | NGC Chromatography Systems and ChromLab SoftwareInserting Check Valves into the Sample Inject ValveAll NGC

Pagina 306 - Laboratories, Inc

Preparing the DetectorsInstrument Guide | 9 74. Connect the tubing to the check valves.Preparing the DetectorsThe NGC systems ship with the conduc

Commenti su questo manuale

Nessun commento