3 | Preparing the Instrument
120 | NGC Chromatography Systems and ChromLab Software
6. Remove the syringe from the priming port and discard the liquid into waste.
7. Repeat steps 2–6 for system pump B.
To purge the system pumps
1. On the ChromLab computer or the touch screen, start ChromLab in Manual
mode.
2. Set the fluidic scheme to either NGC Quest or NGC Quest Plus.
3. In the fluidic scheme, touch or click the Sample Inject Valve module to open its
dialog box.
4. Set the sample inject valve to System Pump Waste.
Vedere la pagina 121