56 | S3 and S3e Cell Sorters
Acquisition
The same can be done for the FL2 positive control, though the compensation required is
much less (Figure 60).
As compensation is adjusted, the compensation matrix will be updated. At any time the
compensation matrix can be viewed using the button on the toolbar (Figure 61).
Fig. 60. Same data plots showing incorrectly compensated file (left) and correctly
compensated file (right).
Fig. 59. Density plot showing uncompensated (left) and compensated data (right).