Instruction Manual | 103
10.4 Swap Nozzle Tip Wizard
38. The AutoGimbal system will commence automatic alignment of the nozzle tip and stream
to the optics.
During the automatic alignment process, a series of windows will appear (Figure 124). This
process can take 56 min to complete.
Fig. 123. Adjust Stream Position window.
Fig. 124. Automated alignment process windows.