iMark Microplate Absorbance Reader Instruction Manual
9
The first time the instrument is turned on, or after a battery failure, the following default information is held in the
memory.
1. Both Administrator and Common user security passwords are set to "00000".
2. Laboratory name is "Bio-Rad Laboratories".
3. All wavelength values of the filters are set to "---" nm.
4. The date and time of the calendar chip are set to 1 day, 1 month, 00 year, 0 hour and 0 minute.
5. No plate reading data is available.
6. End-point protocol is chosen and the protocol number is set to #1.
7. No standard curve data for the storage of standard curves and graphs is available.
Limits
The Microplate Reader displays absorbance readings with absolute values as 3.500. Out-of-range absorbance
values, (those with absolute values greater than 3.500) are displayed as either "*.***" or "-*.***". For example, if
the absorbance is 4.500, then the display will read "*.***", and if the absorbance is -4.500, the display will read
"-*.***".
Vedere la pagina 14