Preparing the System
User Guide | 9 5
5. Prime the system. (See Priming and Purging the Systems in Chapter 3 of the
NGC Chromatography Systems and ChromLab Software Instrument Guide for
detailed instructions.)
To purge the NGC Quest system pumps
1. In the fluidic scheme, touch or click the Sample Inject valve module to open its
dialog box.
2. Set the sample inject valve to System Pump Waste.
3. In the fluidic scheme, double-click the System Pump module to open its dialog
box.
Vedere la pagina 96