System Settings
User Guide | 7 1
3. Select Synchronize with detector.
4. Specify the inner diameter of the tubing in use.
Tubing Color Inner Diameter
Orange 0.02" (0.5 mm)
Green 0.03" (0.75 mm)
Clear 0.062" (1.6 mm)
5. If your system includes the BioFrac fraction collector, select the checkbox
Include BioFrac – 12 l and specify the tubing length between the UV outlet
port and the fraction collector drophead.
6. If your system includes an outlet valve, select the checkbox Include Outlet
valve – 50 l and specify the tubing length between the UV outlet port and the
outlet valve.