ENrich
SEC 70
ENrich
SEC 650
High-Resolution
Size Exclusion Columns
Instruction Manual
Catalog numbers
780-1070
780-1650
Please read these instructions before you use ENrich SEC
high-resolution size exclusion media. If you have any questions or
comments regarding these instructions, please contact your
Bio-Rad Laboratories representative.
Vedere la pagina 0
1 2 3 4 5 6 ... 15 16

Sommario

Pagina 1 - Catalog numbers

ENrich™ SEC 70ENrich™ SEC 650High-Resolution Size Exclusion ColumnsInstruction ManualCatalog numbers780-1070780-1650 Please read these instructions be

Pagina 2

6 ENrich SEC Instruction Manual Section 3: Care of the ENrich SEC Column 3.1 Regular Maintenance To maintain performance, the column should be store

Pagina 3 - Table of Contents

ENrich SEC Instruction Manual 73.3 Column CleaningCareful preparation (especially filtration) of the sample and the buffers will maintain the colu

Pagina 4

8 ENrich SEC Instruction Manual Fig. 2. Column diagram.1. Start a slow flow rate of buffer (0.5 ml/min or less) through the column.2. Remove the lo

Pagina 5 - Size Exclusion Columns

ENrich SEC Instruction Manual 93.5 Storage Conditions To prevent bacterial growth in the column, it must be stored correctly. For short-term stor

Pagina 6 - Chromatography Systems

10 ENrich SEC Instruction Manual Section 4: Ordering Information Catalog # Description780-1070 ENrich SEC 70 10 x 300 Column, 10 x 300 mm780-1650 EN

Pagina 8 - 2.5 Column Calibration

Life ScienceGroupSig 121210027325 Rev A US/EGBio-Rad Laboratories, Inc.Web site www.bio-rad.com USA 800 424 6723 Australia 61 2 9914 2800 Austri

Pagina 10 - : Care of the ENrich SEC

Table of ContentsSection 1: Characteristics of the ENrich Size Exclusion Columns ... 11.1 Introduction ...

Pagina 12 - Fig. 2. Column diagram

ENrich SEC Instruction Manual 1Section 1: Characteristics of the ENrich™ Size Exclusion Columns 1.1 IntroductionENrich prepacked columns for size

Pagina 13 - 3.5 Storage Conditions

2 ENrich SEC Instruction Manual The ENrich support is compatible with aqueous solutions of 6 M guanidine-HCl and 8 M urea. Detergents and organic so

Pagina 14 - : Ordering Information

ENrich SEC Instruction Manual 3Section 2: Use of the ENrich Columns2.1 Preparation for Initial Use The column is supplied in 20% ethanol. To remo

Pagina 15

4 ENrich SEC Instruction Manual 2.3 Sample considerations1. Always filter (0.22–0.45 μm filter) and/or centrifuge your sample to remove any particula

Pagina 16 - Life Science

ENrich SEC Instruction Manual 5Figure 1. Typical chromatograms for ENrich SEC media. Detection at 280 nm. Absorbance at 280 nm may be greater dep